Luminescence 96 plate reader

Using 96 highly sensitive silicon photomultipliers, Luminescence 96 is an entirely new type of luminometer. Its ultra-compact design provides access to cutting-edge technology for every lab workstation ending reliance on centralised equipment. 

Sensitivity

Luminescence 96 is the first plate reader to implement highly sensitive silicon photomultipliers. Sophisticated electronics work to pull large amounts of information coming from these detectors. 

All 96 detectors adapt according to the signal intensity of the corresponding well, reaching a maximum sensitivity of 100 fmol ATP/well.

Reducing phosphorescence

A built in shutter reduces dark count noise and background phosphorescence is cancelled out by a filter.

Dynamic range

Luminescence 96 has a linear dynamic range of up to 8 decades. To access the full potential of the silicon photomultipliers, three readout methods are used: photon counting, integration, and micro integration. 

Cross-talk

With Luminescence 96, every millimetre counts. By brining the microplate withing a 10th of a millimetre of the detection unit, cross-talk is practically eliminated. 

Luminescence 96 App

The Luminescence 96 App is as flexible as the reader itself. It can be used on multiple devices and supports Microsoft Windows and macOS. With plug-and-play technology, simply connect the reader and start your assay. The Luminescence 96 App can be used on a desktop PC in your lab, notebook, or even on Windows tablets. Regular updates bring new features and enhancements, so the user is always up to speed. There are no license limitations and updates are free of charge.

General

Type of product Single-mode luminometer
Detection method Luminescence 
Detection mode  Endpoint, kinetic
Microplate types 96-well microplates, white or black
Readout Flow, Flash (Development times >5 seconds, manual addition of reagents)
Readout orientation  Top and bottom readout 

Measurement 

Detector 96 Silicon Photomultiplier
Sensitivity 100 fmol ATP/well (preliminary result)
Linear dynamic range  8 decades
Cross talk 0.7 x 10⁻⁶  RLU
Well-to-well variability  < 5 %
Data output  USB 2.0 interface with PC or WLAN
Temperature for measurement  15–40 °C 

Physical characteristics 

Dimensions 9,8 x 14,0 x 4,2 cm (W x L x H)
Weight  570 g
Power Supply  USB connection 5 V
Power consumption  2.5 Watts

App 

System requirements 

Microsoft Windows  Windows 7 or higher
macOS Mojave 10.14 or higher

App features 

Measurement modes Endpoint, Kinetic
Evaluation methods Quantitative, qualitative
Regression methods  Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more
Export formats  PDF, CSV
 

Documents

The LabLogic Group will always treat your details with the utmost care and will never sell them to third party companies.

Please choose one of the following:



Previous