A benchtop gamma spectrometer with GMP compliant software

Designed for radionuclidic identity and radionuclidic purity measurements, the Spec-RAM CZT offers a cost-effective alternative to High Purity Germanium (HPGe) detectors, making it the ideal solution for medical radioisotope producers, radiopharmacies, and research facilities when identifying any impurities or contaminants present in a sample.

Improved energy resolution

Cadmium Zinc Telluride (CZT) detectors are highly sensitive to gamma rays – due to the high atomic numbers of Cd and Te – and provide an energy resolution (FWHM) of ≤2.5 compared to traditional scintillator detectors.

Space-saving

With a small footprint and reduced weight, the Spec-RAM CZT can be easily mounted on a lab bench for improved user access unlike HPGe and traditional NaI gamma spectrometers.

The alternative solution to HPGe

The Spec-RAM CZT offsets the disadvantages of larger, costly, more hazardous HPGe systems while still delivering the necessary resolution for the Quality Control measurements of PET and SPECT radiotracers. Perfect for existing facilities which do not have the space, infrastructure, funding, or skillset to operate a HPGe.

Ideal for ambient temperature measurements

CZT detectors can operate in direct-conversion (or photoconductive) mode at room temperature, unlike Germanium which requires cooling.

Energy and efficiency calibration

Laura Radiopharma software includes all the key gamma spectrometry features including both energy and efficiency calibrations.

Excellent signal-to-noise ratio

A copper liner eliminates Compton Scattering, ensuring an excellent signal-to-noise ratio.

Lead shielding

Shields the detector from background environmental radiation. Additionally, it attenuates exposure to the operator.

A compliant solution

As with all LabLogic products, the Spec-RAM CZT has been designed to exceed regulatory requirements. Using our industry standard Laura Radiopharma software, the Spec-RAM CZT ensures regulatory compliance via controlled access and audit trail.

Audit trail

Laura Radiopharma has a fully featured audit trail that can be configured to suit the requirements of the user’s site. Response from the user can be a simple as choosing from a predefined list of standard audit reasons or require a unique response depending on the situation and site rules.

Access levels

Laura Radiopharmacy features multi-level access control, and the system can be configured to have as many levels of access as the user’s site requires. Menu items are then restricted according to the security level of the individual logged in.

Electronic signatures

Laura Radiopharma supports the use of electronic signatures, and the system can be configured to suit the user sites requirements for electronic data.

Data integrity

Regulatory compliance is an essential feature of Laura Radiopharmacy, built to meet GLP/GMP, FDA 21 CFR part 212 / 11, and EU Annex 11 requirements. Featuring configurable audit trail, electronic signatures, secure data storage in a database environment and multi-level security, Laura Radiopharma can be configured to meet regulatory requirements for data integrity.

Security settings

Laura Radiopharma can be configured to suit your particular site requirements for security. Usernames and passwords can be controlled relative to the site requirements as can inactivity time outs and all other functions required for a secure system.

Nuclide ID

Laura Radiopharma’s Nuclide ID function allows users to simultaneously test radionuclidic identification via a half-life measurement and radionuclidic purity using the Spec-RAM CZT. By measuring decay every 15 seconds at the same time as performing gamma spectrometry on one sample, a decay chain is visualised on screen in real-time, providing dozens of datapoints within minutes to automatically identify radionuclides.

 